The HC-i is designated as DCE (Data Communication Equipment).

13-3. Data Output Mode

‰

The Data Output Modes and Parameters are set by F-Functions in F-06-X as

described in the “9-2. F-Functions”

‰

To control the scale using commands form an external device, see “13-6. Command

Mode”.

‰

Refer to “13-6. Command Mode” about the output data format.

Data Output Mode (f-06-01)

‰

Key Mode (f-06-01="0")

When the weight display is stable, data is sent by pressing the

PRINT

key. The

count display will blink when the data has been sent.

‰

Stream Mode (f-06-01="1")

Data is sent continuously. The data-update rate is approximately 10 times per second for
f-06-03="2". For f-06-3="0" or "1", the interval between continuous data is
approximately 2 seconds.

‰

Auto-print Mode A (f-06-01="2")

Data is sent if the weight display is stable at +5d (weighing display division) and
above. The next transmission can not occur until after the weight display falls below
+5d.

‰

Auto-print Mode B (f-06-01="3")

Data is sent if the weight display is stable at ±5d (weighing display division) and
above/below. The next transmission can not occur until after the weight display falls
between –5d and +5d.
To use with the UFC format, refer to "13-7. Using UFC (Universal Flex Coms)
Function"

Data to be Sent (f-06-02)

Select which data to be sent by keying in a

0

or

1

for the data: ID No., PCS

(count), weight or unit weight.

Example: Key in

1

1

0

0

to

display 1100, this setting would
send only the ID number and the
count.





Data Format (f-06-03)

‰

Format for AD-8121 MODE 1 or 2. (f-06-03="0")

‰

Format for AD-8121 MODE 3. (f-06-03="1")

‰

Format for general apparatuses, computers, etc. (f-06-03="2")

‘0’ not to send data
‘1’ to send data

8

f-06-02

88Clear8

8

ai

1100

0

00

ID No.

PCS (count)

subtr

acts