11hi00 – A&D Weighing HC-3Ki User Manual
Page 31
29
8-2. Gravity Compensation
When the scale is first used or has been moved to different place, it should be
calibrated using a calibration weight.
But if the calibration weight cannot be prepared, the gravity acceleration correction
will compensate the scale. Change the gravity acceleration value of the scale to the
value of the area where it will be used. Refer to the gravity acceleration map
appended to the end of this manual.
1. In the Step 1 above, press
TARE
key
The display flashes the gravity acceleration
value stored in the scale.
2. Use the
0
Æ
9
10-key pad to
display the desired gravity
acceleration value.
(Example of the value 9.800 m/s
2
.)
3. Press the
ENTER
key.
The scale will store the new value.
If necessary to calibrate the scale
using a weight, go to Step 2 of the
previous section.
4. Press the
ON/OFF
key to turn the scale off and re-attach the calibration switch
cover. (End of the calibration procedure.)
ENTER
888
Cal
10
COUNT
WEIGHT UNIT WEIGHT
88
g
al008
8
0.0
9.798
0
00
888
Cal
10
COUNT
WEIGHT UNIT WEIGHT
88
g
al008
8
0.0
9.800
0
00
888
Cal
02
COUNT
WEIGHT UNIT WEIGHT
88-
0.000
8
8
11Hi00