Data format – A&D Weighing HC-3Ki User Manual
Page 50
48
If the command is
S
,
T
or
Z
, the scale will send a second acknowledgment
R > F > or operation is completed. If an error occurs, the scale will send an error code. The error format is E C , E n C L F , “n” being error number. En Definition Notes E0 Communication Error Parity error, framing error, etc. E1 Undefined command Error Command does not exist for the scale. E2 Scale not ready Error. The scale is not in a state where a command E4 Too many characters Error Command contains too many characters. E6 Format Error Command contains invalid characters. E7 Out of range Error Value is out of range. Tare weight is more than Data Format Examples below are for f-09-02=”0000”. Store unit weight and other value in use (according to f-01-05). Command E , 1 2 C L F Stores to id-12. (E,000012C R L F is acceptable.) Reply ACK C L F ID Number Command ? I D C L F Reply I D , 0 0 0 0 1 2 C L F PCS (Count ) Data Command ? Q T C L F Reply Q T , + 0 0 0 0 1 2 3 4 [ P C C L F Stable Positive Data U S , - 0 0 0 0 5 6 7 8 [ P C C L F Unstable Negative Data O L , + 9 9 9 9 9 9 9 9 [ P C C L F ‘E’ display Weight Data Command ? W T C L F Reply S T , + 0 0 1 . 2 3 4 6 [ k g C L F Stable Positive Data S T , - 0 0 2 . 7 2 5 5 [ l b C L F Stable Negative Data U S , - 0 0 1 2 . 3 4 6 [ l b C L F Unstable Negative Data U S , + 0 0 0 5 . 5 9 3 [ k g C L F Unstable Positive Data O L , + 9 9 9 9 . 9 9 9 [ k g C L F ‘E’ display O L , - 9 9 9 9 . 9 9 9 [ l b C L F ‘-E’ display “ [ ” in examples below shows “Space” (20H).
could be expected.
the capacity, etc.