If the command is

S

,

T

or

Z

, the scale will send a second acknowledgment

R

>

F

> or (see F-Function “f-09-02”) when the command

operation is completed.

If an error occurs, the scale will send an error code.

‰

The error format is

E C , E n C L

F

, “n” being error number.

En Definition

Notes

E0

Communication Error

Parity error, framing error, etc.

E1

Undefined command Error Command does not exist for the scale.

E2

Scale not ready Error.

The scale is not in a state where a command
could be expected.

E4

Too many characters Error Command contains too many characters.

E6

Format Error

Command contains invalid characters.

E7

Out of range Error

Value is out of range. Tare weight is more than
the capacity, etc.

Data Format

‰

Examples below are for f-09-02=”0000”. =06H.

‰

Store unit weight and other value in use (according to f-01-05).

Command E , 1 2 C L

F

Stores to id-12. (E,000012C

R

L

F

is acceptable.)

Reply ACK C L

F

‰

ID Number

Command ? I D

C L

F

Reply

I D , 0 0 0 0 1 2 C L

F


‰

PCS (Count ) Data

Command ?

Q T

C L

F

Reply

Q T , + 0 0 0 0 1 2 3 4

[ P C C L

F

Stable Positive Data

U S , - 0 0 0 0 5 6 7 8

[ P C C L

F

Unstable Negative Data

O L , + 9 9 9 9 9 9 9 9

[ P C C L

F

‘E’ display


‰

Weight Data

Command ?

W T

C L

F

Reply

S T , + 0 0 1 . 2 3 4 6

[ k g C L

F

Stable Positive Data

S T , - 0 0 2 . 7 2 5 5

[ l b C L

F

Stable Negative Data

U S , - 0 0 1 2 . 3 4 6

[ l b C L

F

Unstable Negative Data

U S , + 0 0 0 5 . 5 9 3

[ k g C L

F

Unstable Positive Data

O L , + 9 9 9 9 . 9 9 9

[ k g C L

F

‘E’ display

O L , - 9 9 9 9 . 9 9 9

[ l b C L

F

‘-E’ display

[

” in examples below shows “Space” (20H).