Specifications – A&D Weighing GX-8000 User Manual
Page 84
82
21. SPECIFICATIONS
GX-200
GX-300
GX-400
GX-600
GX-800 GX-1000
Weighing capacity
210 g
310 g
410 g
610 g
810 g
1100 g
Maximum display
210.084 g 310.084 g 410.084 g 610.084 g 810.084 g 1100.084 g
Minimum weighing value (1 digit)
0.001 g
Repeatability (Standard deviation)
0.001 g
Linearity
±0.002 g
±0.003 g
Stabilization time (typical at FAST)
Approx. 1 second
Approx. 1.5 seconds
Sensitivity drift
(10°C-30°C/50
°F-86°F )
±2 ppm/°C (When automatic self calibration is not used)
±0.010 g
Accuracy right after calibration using
the internal mass. See notes below
The value above is to the weighing capacity.
Operating environment
5°C to 40°C (41°F to 104°F)
85%RH or less (No condensation)
Display refresh rate
5 times/second or 10 times/second
Minimum unit mass
0.001 g
Counting
mode
Number of samples
10, 25, 50 or 100 pieces
Minimum 100% reference mass
0.100 g
0.001 g
Percent
mode
Minimum 100% display
0.01%, 0.1%, 1% (Depends on the reference mass stored.)
External calibration weight
200 g
100 g
300 g
200 g
100 g
400 g
300 g
200 g
600 g
500 g
400 g
300 g
200 g
800 g
700 g
600 g
500 g
400 g
300 g
200 g
1000 g
900 g
800 g
700 g
600 g
500 g
400 g
300 g
200 g
Weighing pan
128 x 128 mm
Net weight
Approx. 4.6 kg
External dimensions
210 (W) x 317 (D) x 86 (H) mm
AC adapter
Confirm that the adapter type is correct for the local voltage and power
receptacle type
Power consumption
Approx. 11VA (supplied to the AC adapter)
Interface (Provided as standard)
RS-232C with Windows Communication Tools WinCT
Notes
Accuracy right after calibration using the internal mass in good ambient conditions (within
the temperature range of 10
°C to 30°C (50°F to 86°F) with no abrupt changes in temperature
or humidity, no drafts, no effect by magnetic fields or static electricity).
The internal mass has a mass of approximately 500 grams.