beautypg.com

A&D Weighing GX-8000 User Manual

Page 41

background image

39

Zero tracking ( trc )

This function tracks zero point drift caused by changes in the environment and stabilizes the zero
point. When zero has difficulty stabilizing, select a greater value. When the weighing data is only a
few digits, turn the function off for accurate weighing.

trc 0

The tracking function is not used. Used for weighing a very light sample.

trc 1

The tracking function is used. Normal zero tracking.

trc 2

The tracking function is used. Strong zero tracking.

trc 3

The tracking function is used. Very strong zero tracking.

Note
Digit, when used for the GX series balance, indicates a unit of minimum weighing value.

Display refresh rate ( 5pd )

Period to refresh the display. This parameter influences “Baud rate”, “Data output pause” and
“Stream mode”.

Note
This item is selected automatically in the automatic response adjustment.

Decimal point ( pnt )

The decimal point format can be selected.

Auto display-ON ( p-on )

When the AC adapter is connected, the display is automatically turned on without the ON:OFF key
operation, to display the weighing mode. Used when the balance is built into an automated system.
30-minute warm up is necessary for accurate weighing.

Auto display-OFF ( poff )

When the AC adapter is connected and no operation is performed (inactivity state) for 10 minutes,
the display is automatically turned off and the standby indicator is illuminated.

Capacity indicator ( g5i )

In the weighing mode, the indicator displays the weighing data relative to the weighing capacity in
percentage. (Zero = 0%, maximum capacity = 100%)
When the “Data memory ( data )” parameter is set to “1” (to store unit mass in the counting mode) or
to “2” (to store the weighing data), the indicator displays the information stored in memory, such as
the amount of memory data or data number.

This manual is related to the following products: