A&D Weighing GX-8000 User Manual
Page 33
31
4 Press the PRINT key. Then the balance displays the
function switches.
Set the function table switch and internal mass correction
switch to “1” as shown above using the following keys.
SAMPLE key
To select the switch to change the value.
RE-ZERO key To change the parameter of the switch
selected.
5 Press the PRINT key to store the new setting. The
balance returns to the weighing mode.
6 Press and hold the SAMPLE key to enter the function
table and release the key when ba5fnc is displayed.
7 Press the SAMPLE key several times until C5in is
displayed.
8 Press the PRINT key to enter the procedure for correcting
the internal mass value.
9 Correct the internal mass value using the following keys.
RE-ZERO key To select the value.
(-20 digits appear after +20 digits.)
PRINT key
To store the new value and display the
next menu of the function table.
CAL key
To cancel the correction and display
the next menu of the function table.
10 Press the CAL key. The balance returns to the weighing mode.
11 Press the CAL key to calibrate the balance using the
internal mass.
12 Place the external weight on the pan and confirm that the correction has been performed
properly. In this example, confirm that the value displayed is within
±2 digits of 200.000 grams,
or within
±0.002 gram.
If the internal mass value has not been corrected properly, repeat the procedure, starting at
step 6 and adjust the correction value.
Press and