6 description of the item “data format – A&D Weighing GX-8000 User Manual
Page 43
41
Interval memory mode
The weighing data is periodically stored in memory.
Required setting
dout prt
3
Interval memory mode
dout data
2
Data memory function is used.
Stores weighing data.
dout int Interval
time
Optional setting
dout
5-td1, 2, or 3 Adds the time and date.
Example
“For periodical weighing without a computer command and
outputting all of the data to a computer at one time”
10-6 Description of the Item “Data format”
A&D standard format
5if type 0
This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B is
used, set the printer to MODE 1 or 2.
This format consists of fifteen characters excluding the terminator.
A header of two characters indicates the balance condition.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus
sign is used.
The unit, consisting of three characters, follows the data.
DP (Dump print) format
5if type 1
This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B
is used, set the printer to MODE 3.
This format consists of sixteen characters excluding the terminator.
A header of two characters indicates the balance condition. No overload header is used.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is
not zero or overloaded.
The unit, consisting of three characters, follows the data.
Unit
Data
Header
S T , + 0 0 0 1 2 . 7 8
8 8
g
CR LF
Terminator