beautypg.com

Precautions, 1 before use – A&D Weighing GX-8000 User Manual

Page 10

background image

8

3. PRECAUTIONS

To get the optimum performance from the balance and acquire accurate weighing data, note the

following:

3-1 Before Use

Install the balance in an environment where the temperature and

humidity are not excessive. The best operating temperature is
about 20°C / 68°F at about 50% relative humidity.

Install the balance where it is not exposed to direct sunlight and

it is not affected by heaters or air conditioners.

Install the balance where it is free of dust.
Install the balance away from equipment which produces

magnetic fields.

Install the balance in a stable place avoiding vibration and

shock. Corners of rooms on the first floor are best, as they are
less prone to vibration.

The weighing table should be solid and free from vibration,

drafts and as level as possible.

Level the balance by adjusting the leveling feet and confirm it

using the bubble spirit level.

Ensure a stable power source when using the AC adapter.
Warm up the balance for at least 30 minutes. Plug in the AC

adapter as usual.

Calibrate the balance before use or after having moved it to

another location.

Caution

Do not install the balance where flammable or corrosive gas is present.

This manual is related to the following products: