beautypg.com

A&D Weighing GX-8000 User Manual

Page 39

background image

37

Caution

The balance may not transmit the data completely at the specified refresh rate, depending on
the baud rate or data added to the weighing data such as time, date and ID number.

Class

Item

P

aram-

eter

Description

Serial interface

Density function

Programmable-unit (Multi-unit)

Unit

Internal mass value correction

ID number setting

Baud rate

Data bit, parity bit

Terminator

Data format

Timeout

AK, Error code

CTS, RTS control

Liquid density input

5if

d5 fnc

mlt

Unit

C5 in

id

bp5

btpr

Crlf

type

t-Up

erCd

Ct5

ldin

0
1
2
3
4
0
1
2
0
1
0
1
2
3
4
5
0
1
0
1
0
1

0

1

*

600 bps

1200 bps
2400 bps
4800 bps
9600 bps
7 bits, even
7 bits, odd
8 bits, none
CR LF
CR
A&D standard format
DP format
KF format
MT format
NU format
CSV format
No limit
1 second
No output
Output
Not used
Used

Water temperature

Liquid density

CR: ASCII code 0Dh
LF: ASCII code 0Ah

See "10-6 Description of
Item "Data Format".

Selects the wait time to
receive a command.
AK: ASCII code 06h

Controls CTS and RTS.

Available only when
density mode is selected.
See "15. DENSITY
MEASUREMENT".
Available only when

programmable-unit mode

is selected.

Displayed only when the

internal mass value

correction switch is set to 1.

See "8. CALIBRATION".

See "11. ID NUMBER AND GLP REPORT".

Factory setting

Sets an arbitrary coefficient.
See "14. PROGRAMMABLE-UNIT".

See "5. WEIGHING UNITS".

This manual is related to the following products: