A&D Weighing GX-8000 User Manual
Page 81
79
Display Error code
Description
EC, E20
Calibration weight error
The calibration weight is too heavy. Confirm the
calibration weight value.
Press the CAL key to return to the weighing mode.
EC, E21
Calibration weight error
The calibration weight is too light. Confirm the
calibration weight value.
Press the CAL key to return to the weighing mode.
Overload error
A sample beyond the balance weighing capacity has
been placed on the pan.
Remove the sample from the pan.
Weighing pan Error
The weight value is too light.
Confirm that the weighing pan is properly installed and
calibrate the balance.
Sample mass error
The balance can not store the sample for the counting
mode or for the percent mode because it is too light.
Use a larger sample.
Unit mass error
The sample unit mass for the counting mode is too
light. Storing and using it for counting will cause a
counting error.
Add samples to reach the specified number and press
the PRINT key.
Pressing the PRINT key without adding samples will
shift the balance to the counting mode. But, for
accurate counting, be sure to add samples.
Automatic response adjustment zero error
The automatic response adjustment can not be
performed because there is something on the pan.
Clear the pan. Press the CAL key to return to the
weighing mode.
(CHECK NG)
Automatic response adjustment unstable error
The automatic response adjustment can not be
performed because the weight value is unstable.
Check the ambient conditions such as breeze,
vibration and magnetic fields, also check the weighing
pan. Press the CAL key to return to the weighing mode.
(CHECK NO)
Internal error
Indicates an internal error as the result of self-check
function.
Repair is required. Contact the local A&D dealer.