A&D Weighing GX-8000 User Manual
Page 51
49
9
With Cp lnt displayed, press the PRINT key. The current setting is displayed. When the
current setting is to be changed, change the setting using the following keys. Enter the
tolerance value in percentage to the reference value as 100%.
SAMPLE key
To select the digit to change the value.
RE-ZERO key
To change the value of the digit selected.
PRINT key
To store the new setting and go to step 10.
CAL key
To cancel the new setting and go to step 10.
10 Press the CAL key to exit the comparator function and return to the weighing mode.
Setting example 2
(Comparison when the weighing data is stable or overloaded, including “near zero”, upper limit and
lower limit, weighing input)
Selecting a comparator mode
1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed.
2 Press the SAMPLE key several times to display Cp fnc .
3 Press the PRINT key.
4 Press the RE-ZERO key several times to display Cp 2 .
5 Press the SAMPLE key several times to display Cp in .
6 Press the RE-ZERO key several times to display Cp in 1 .
7 Press the PRINT key to store the selected mode.
Entering the upper and lower limit values
8 With Cp Hi displayed, press the PRINT key. The current setting is displayed with all of the
digits blinking. Press the RE-ZERO key to enter the weighing input mode.
9 Press the RE-ZERO key. The balance displays 0.00 g. Place a sample whose mass
corresponds to the upper limit value on the pan. Press the PRINT key to store the upper limit
value. Remove the sample. The balance displays Cp lo .
10 With Cp lo displayed, press the PRINT key. The current setting is displayed with all of the
digits blinking. Press the RE-ZERO key to enter the weighing input mode.
11 Press the RE-ZERO key. The balance displays 0.00 g. Place a sample whose mass
corresponds to the lower limit value on the pan. Press the PRINT key to store the lower limit
value. Remove the sample.
12 Press the CAL key to exit the comparator function and return to the weighing mode.
Notes
When Pound/Ounce is selected as a weighing unit, enter the values in ounces for
comparison.
In the density mode, comparison is performed to the density obtained.