9 clock and calendar function – A&D Weighing GX-8000 User Manual
Page 48
46
10-9 Clock and Calendar Function
The balance is equipped with a clock and calendar function. When the “GLP output (info)”
parameter is set to “1” or “2” and the “Time/Date output (5-td)” parameter is set to “1”, “2” or “3”, the
time and date are added to the output data. Set or confirm the time and date as follows:
Operation
1 Press and hold the SAMPLE key until ba5fnc of
the function table is displayed.
2 Press the SAMPLE key several times to display
Cl adj .
3 Press the PRINT key. The balance enters the mode to
confirm or set the time and date.
Confirming the time
4 The current time is displayed with all the digits
blinking.
When the time is correct and the date does not
need to be confirmed, press the CAL key and
proceed to step 8.
When the time is correct and the date is to be
confirmed, press the SAMPLE key and proceed to
step 6.
When the time is not correct and is to be changed,
press the RE-ZERO key and proceed to step 5.
Setting the time (with part of the digits blinking)
5 Set the time in 24-hour format using the following
keys.
SAMPLE key
To select the digits to change
the value. The selected digits
blink.
RE-ZERO key
To increase the value by one.
MODE key
To decrease the value by one.
key
To store the new setting,
display end and go to step
6.
CAL key
To cancel the new setting and
go to step 6.