The relation between the water temperature and density is shown below.
Entering the density directly ( ldin 1 )
In the density mode, press and hold the MODE key until the density currently set (unit : g / cm
3
,
factory setting : 1.000g / cm
3
) is displayed. Use the following keys to change the value.
RE-ZERO key
To set the value of the digit selected.
SAMPLE key
To select the digit to change the value.
PRINT key
To store the change, display end and return to the density mode.
CAL key
To cancel the change and return to the density mode.
Note
The range to set the density is 0.0000 to 1.9999 g / cm3
Measuring the density
In the density measurement, the balance displays the weight of a sample in air, the weight in liquid
and then the density.
Measuring the weight of a sample in air.
The processing indicator blinks with the unit “g” displayed.
Measuring the weight of a sample in liquid.
The processing indicator illuminates with the unit “g” displayed.
Displaying the density.
The processing indicator illuminates with “DS” displayed.
To cycle through the above three, press the SAMPLE key.
Note
The SAMPLE key can not be used to change the minimum weighing value. The density is
displayed with three decimal places.