A&D Weighing GX-8000 User Manual
Page 45
43
10-7
Description of the Data Format Added to the Weighing Data
Data number dout d-no 1
This format outputs the data number just before the data is transmitted using the RS-232C interface.
This format consists of six characters excluding the terminator.
When CSV format (5if type 5) is selected, the period (
. ) is replaced with a comma (
,
).
Note
The data number is added only to the weighing data that is stored in memory.
Time dout 5-td 1
or
3
Outputs time in 24-hour format.
Date dout 5-td 2
or
3
The date output order can be changed in “Clock (Cl adj)”. Outputs the year in four-digit
format.
ID number
dout 5-id 1
The number to identify a specific balance.
This format consists of seven characters excluding the terminator.
Note
When the data described above is added to the weighing data, the output is in the following
order: ID number, Data number, Date, Time and Weighing data.