The following commands can not be used during data storage.
Q
Query command for weighing data.
S
Query command for stable weighing data.
SI
Query command for weighing data.
SIR
Query command for continuous weighing data.
Setting the function table
Parameter settings for each output mode are as follows:
Item
Mode
Data output
mode
Auto print
polarity,
difference
Data memory
function
Interval
time
Key mode
prt 0
Not used
Data 2
Auto print mode A
prt 1
Data 2
Auto print mode B
prt 2
ap-a 0-2
ap-b 0-2
Data 2
Not used
Interval memory mode
prt 3
Not used
Data 2
int 0-8
Set each item, depending on the situation, as follows:
No
d-no 0
No
5-td 0
Data number
Yes
d-no 1
Time only
5-td 1
No
5-id 0
Date only
5-td 2
ID number
Yes
5-id 1
Time and date
Both
5-td 3
With 5-td 1, 5-td 2 or 5-td 3 selected, the amount of data to be stored is 100 sets.
Enabling the data memory function
1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed.
2 Press the SAMPLE key several times to display dout .
3 Press the PRINT key.
4 Press the SAMPLE key three times to display data 0 .
5 Press the RE-ZERO key to change data 0 to data 2 .
6 Press the PRINT key to save.
7 Press the CAL key to return to the weighing mode.
Recalling the memory data
Confirm that the “Data memory ( data )” parameter is set to “2”.
1 Press and hold the PRINT key until recall is displayed.
2 Press the PRINT key to enter the memory recall mode. Recall the data in memory using the following keys
RE-ZERO key
To proceed to the next data set.
MODE key
To go back to the previous data set.
PRINT key
To transmit the current data using the RS-232C interface.
With SAMPLE held down, press the CAL key
To delete the data currently displayed.
Note: Deleting the data will not increase the number of data that can be stored.
CAL key
To exit the memory recall mode.
3 Press the CAL key to return to the weighing mode.