Cal Low Command
(CALLOWCMD)
The Cal Low Command - sets the “calLowCount”
parameter to the current A/D average counts when
doing a hard calibration.
Error Return Values:
•
STATUSWORD - there was a conversion error,
weight in motion or an A/D error or all three.
•
HARDCALFAILCOUNTS - there is less than
100 counts between the zero and the span
weights.
SEQ
NUMBER
0
COMMAND#
(Hex)
1
STATUS
2
OFFSET
3
LENGTH
4
DATA
5-16
From M1 File
0x64
Cal High
Command
(CALHIGHCMD)
The Cal High Command - Sets the Span Weight
parameter to the current A/D average counts when
doing Hard CAL.
Error Return Values:
•
STATUSWORD - there was a conversion error,
weight in motion or an A/D error or all three.
•
HARDCALFAILCOUNTS - there is less than
100 counts between the zero and the span
weights.
SEQ
NUMBER
0
COMMAND#
(Hex)
1
STATUS
2
OFFSET
3
LENGTH
4
DATA
5-16
From M1 File
0x65
C2 Cal Command
(C2CALCMD)
The C2 Cal Command - Performs a C2 Calibration.