Page 40
2.
Attach one end of the connection hose (28) with the rubber sleeve
(29) to the adapter (27) and the other end to the cyclone sepa-
rator (30).
NOTICE!
There are 2 opposing hose connections on the
cyclone separator (30). Do not use the sealed con-
nection; leave it closed. This would increase the noise
level!
3.
Close the cutting mill with mounted standard funnel (1) (see
Chapter 6.1.6 ‘Closing the cutting mill’ on page 34).
4.
Screw the sample bottle (31) onto the adapter (33) on the bottom
of the cyclone separator (30); make sure the connection hose is
firmly connected and switch on the exhaust system.
5.
Switch on the cutting mill (see Ä Chapter 5.2 ‘Switching on’
6.
Add a little comminution material by hand and observe the oper-
ating noises. If the motor speed decreases audibly, reduce the
supply of comminution material. A fast vortex of comminution
material must form in the collection bottle.
Using the device