6.3.3.2

Transmission frequency setting options

Transmission every

Switch 3

Switch 4

1 second

OFF

OFF

½ second

ON

OFF

¼ second

OFF

ON

automatic

ON

ON

1-second transmission is the default value.
For automatic transmission, the 1 / ½ / ¼ second transmission frequency

is switched over in the event of fast changes in pressure or temperature.
The data transmission frequency naturally also has an impact on the bat-

tery's service life.
1-second operation results in the longest time span.
The transmission frequency can also be checked visually: as soon as the

transmission unit is switched on, the switch flashes each time data is

transmitted.

6.3.4 Installation of the transmission unit on the EASY GTM bowl

NOTICE!
The threaded holes inside the bowl, the passage holes and

the Allen screws in the locking ring have to be cleaned thor-

oughly after each grinding to allow a safe fastening of the

EASY GTM system.

The EASY GTM system must be assembled as follows before inserting the

EASY GTM system in the micro mill:

Accessories

- 75 -