6.2 Gassing lid with stainless steel, swagelok hose couplings
NOTICE!
Our workers perform a water bath test on all gassing lids.
The part to be tested is sealed, a pressure of 5.5 bar is
applied and it is immersed in a water bath. If there is a leak,
bubbles will develop. The air bubbles that develop within a
specific interval are evaluated by the worker/tester.
Only gassing lids with a leak rate of <10
-4
[mbar l/s] are
approved.
6.2.1 Scope of delivery / lid design
1 Compl. gassing lid handle
2 Compl. gassing lid fitting
3 Gassing valve connecting plug
4 Gassing valve quick-release coupling
5 Sealing ring 8x12x1mm, vulcanised fibre
6 Flat seal
7 Actuating pin
8 Gassing lid insert
9 Flat seal, 57.5x48x2, silicone
10 Hexagon socket wrench size 13
Accessories
- 66 -