Checkline DFX-7 User Manual
Page 24
Dakota Ultrasonics
20
purpose of the design, was to provide the user with an efficient way to make
adjustments on the fly during inspections. The following procedure outlines the steps
to navigate and make adjustments as follows:
Hot Menu
Gate1
1) Press the MEAS key once to activate the Hot Menu. Press the MEAS key
multiple times to advance the cell right, and the ESC key multiple times to
advance the cell left, until the WIDTH1 cell is highlighted.
2) Press the LEFT and RIGHT arrow keys to toggle the value or status.
Note: This is a coarse adjustment. It’s the fastest and easiest way to
increment, decrement, or toggle the status, however it does so in coarse steps.
Set Detect
1) Press the UP and DOWN arrow keys until DETECT is highlighted.
2) Press the LEFT and RIGHT arrow keys to toggle the ABOVE or BELOW.
Note: Above will trigger the alarm when a signal meets or exceeds the height
of the gate, and below will detect on anything that meets or is below the height
of the gate.
3) Alternatively, press the ENTER key to display the Digits Edit Box.
4) Press the UP and DOWN arrow keys to scroll the highlighted value.
5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.