Brookfield YR-1 Rheometer User Manual
Page 6

Brookfield Engineering Labs., Inc.
Page 6
Manual No. M02-215-B0412
spindle. The maximum torque value is the yield point. The straight line in Figure I-3 is a repeat
of what was shown in Figure I-2. An algorithm in the YR-1 firmware converts the maximum
torque value into a yield stress value.
The shear stress measurement range of the YR-1 (in Pascals) is determined by the size and shape
of the vane spindle, and the full scale torque range of the calibrated spring.
There are four basic spring torque series offered by Brookfield:
Spring Torque
Model
dyne-cm
milli Newton - m
1/4RVYR-1 1,796 0.1796
RVYR-1
7,187 0.7187
HBYR-1
57,496 5.7496
5XHBYR-1 287,480 28.7480
The shear stress measurement range for the three standard (or supplied) vane spindles at each
spring torque is as follows:
SPINDLE
TORQUE RANGE
SHEAR STRESS
RANGE (Pa)
V-71
1/4 RV
.125-1.25
V-72
1/4 RV
.5-5
V-73
1/4 RV
2.5-25
V-71
RV
.5-5
V-72
RV
2-20
V-73
RV
10-100
V-71
HB
4-40
V-72
HB
16-160
V-73
HB
80-800
V-71
5xHB
20-200
V-72
5xHB
80-800
V-73
5xHB
400-4000
Table I-1
The equivalent units of measurement in the SI system are calculated using the following con-
versions:
SI
CGS
Shear Stress: 1 Pascal or 1 Newton/m
2
=
10 dyne/cm
2
Torque:
1 Newton-m
=
10
7
dyne-cm