Appendix c - calibration procedures – Brookfield YR-1 Rheometer User Manual
Page 46
Brookfield Engineering Labs., Inc.
Page 46
Manual No. M02-215-B0412
APPENDIX C - Calibration Procedures
For more help you can go to the website,
www.brookfieldengineering.com, and download the video.
The torque sensing spiral spring in the YR-1 Rheometer is calibrated at Brookfield to be accu-
rate within ±1% of full scale range of a certified master viscometer. However, when checking
calibration with a vane spindle using the following procedures, an accuracy of ±2% of full scale
range is allowed.
The accuracy of the YR-1 is verified using one of your vane spindles with the appropriate viscos-
ity standard fluids which are available from Brookfield Engineering Laboratories or your local
Brookfield agent. Viscosity standards are Newtonian and, therefore, have the same viscosity
regardless of spindle speed. Viscosity standards, calibrated at 25°C, are shown in Table C-1
(Silicone Oils) and Table C-2 (Mineral Oils).
GENERAL PURPOSE SILICONE FLUIDS
BEL Part No.
Viscosity (cP) 25ºC
5 CPS
5
10 CPS
10
50 CPS
50
100 CPS
100
500 CPS
500
1,000 CPS
1,000
5,000 CPS
5,000
12,500 CPS
12,500
30,000 CPS
30,000
60,000 CPS
60,000
100,000 CPS
100,000
Table C-1 - Silicone Viscosity Standard Fluids
MINERAL OIL VISCOSITY STANDARD FLUIDS
BEL Part No.
Viscosity (cP) 25ºC
B29
29
B200
200
B600
600
B1060
1,060
B2000
2,000
B10200
10,200
B21000
21,000
B730000
73,000
B200000
200,000
B360000
360,000
Table C-2 - Mineral Oil Viscosity Standard fluids