Brookfield YR-1 Rheometer User Manual
Page 34
Brookfield Engineering Labs., Inc.
Page 34
Manual No. M02-215-B0412
III.4.2
Post-Test Output
At the conclusion of the test, one of the following six (6) messages is sent to both the parallel
and serial (RS-232) ports:
1. Test (test number: test name) Complete
Yield Stress (Pa) = 196.53
% Torque @ Yield = 78.6 Temperature = 25.5C
Test Passed
2. Test (test number: test name) Complete
Test Failed = Under-range (displayed if test fails due to an under-range condition; see Section III.6.2)
3. Test (test number: test name) Complete
Test Failed = Over-range
(displayed if test fails due to an over-range condition; see Section III.6.2)
4. Test (test number: test name) Complete
Yield Stress (Pa) = 196.53
% Torque @ Yield = 78.6 Temperature = 25.5C
Test Failed = Yield Stress Below Low Limit
(displayed if test fails due to resultant yield stress
being below low limit; see Section III.6.2)
5. Test (test number: test name) Complete
Yield Stress (Pa) = 196.53
% Torque @ Yield = 78.6 Temperature = 25.5C
Test Failed = Yield Stress Above High Limit
(displayed if test fails due to resultant yield stress
being above high limit; see Section III.6.2)
6. Test (test number: test name) Cancelled
Test Failed = Cancelled By User (displayed if test fails due to cancellation by the user)
III.5 Printing a Program
After selecting PRINT TEST (by pressing the Enter key with PRINT TEST flashing on the
bottom line) from the default screen (Figure III-6), the following is displayed:
#X yyyyyyyyyy
↑↓
PRINT TEST
Figure III-19
Note: While the top line is the active selection line, the program number and name will
be flashing and the ↓ and ↑ keys will be displayed to the right of the program name.
Pressing the ↓ or ↑ key scrolls through the program numbers (0-9) and displays the correspond-
ing program name. While scrolling, if the memory location scrolled to has no program in it,
EMPTY
is displayed and pressing the Enter key has no effect.
Pressing the ESC key with this screen displayed causes the screen in Figure III-6 to again be
displayed (i.e. go back one level).
Pressing the Enter key with a valid program selected displays the following screen: