Brookfield YR-1 Rheometer User Manual
Page 32
Brookfield Engineering Labs., Inc.
Page 32
Manual No. M02-215-B0412
TEST #5 SPDL 71
RUNNING YIELD
Figure III-13
If the eSC key is pressed at any time during the yield test run, the following is displayed:
TEST TERMINATED
BY USER
Figure III-14
At this point, pressing the Enter key starts the test again starting from whichever Run screen
is applicable (Figures III-10 through III-13). Pressing the ESC key returns the display to that
of Figure III-8, reminding you which spindle should be used.
When the test is allowed to run to completion, one of the following screens (Figures III-15
through III-18) are displayed:
Figure III-15 is displayed when the yield test passes. The % torque and stress (in Pa) at the
point of yield are displayed along with the temperature (in °C) at the conclusion of test.
%=73.56
025.5C
YIELD=726.36
Pa
Figure III-15
Figure III-16 is displayed when the yield test fails due to an over-range error (see Section
III.6.2).
YIELD ERROR
OVER-RANGE
Figure III-16
Figure III-17 is displayed when the yield test fails due to an under-range error (see Section III.6.2).
YIELD ERROR
UNDER-RANGE
Figure III-17