Troubleshooting

This section presents tables of information to help you troubleshoot issues on the

monitor.

To use these tables, locate the specific problem with your central station in the left

column of the table. The remainder of the row explains possible causes and suggests

actions that can resolve the issue.

Display

Issue

Possible cause

Suggested action

The central station display is blank The cable for the display is not connected

Connect the monitor display

cables.

The display is not powered on

Power on the display.

Audio

Issue

Possible cause

Suggested action

No audible alarm tones are coming

from the central station display

The display is muted

Look for a mute button on the

display and press it to restore

sound.

The audio cables are disconnected

Connect the audio cables.

Headphones are connected to the PC headphone

jack

Disconnect the headphones.