TM 262

7-6

Revised 4/11/08

Intensity Levels

125 Hz

-10 to 50 dB HL

500 to 6000 Hz-10 to 90 dB HL

250 and 8000 Hz-10 to 70 dB HL

NOTE: An additional +10 dB is available per frequency via
the +10 dB pushbutton.

Accuracy:

125 to 4000 Hz ±3 dB

6000 and 8000 Hz ±5 dB

Step Size:

5 dB

Signal-to-Noise Ratio:

> 70 dB in 1/3 octave; less than -10 dB HL for levels less than
60 dB HL

Rise/Fall Time:

20 to 50 msec

Tone Format:

Tone is normally off until present bar is depressed.

Continuous

Steady when present bar is depressed

Pulsed

2.5/sec (i.e., 200 msec ON, 200 msec OFF)

FM (frequency modulated)

5 Hz, ±5%

Transducers

IPSI:

Welch Allyn design

Audiometric Headset:

Pair TDH-39 earphones with MX41AR cushions

(60 ohms impedance)

Headband force per ANSI S3.6 and IEC 645 (4.5 ±0.5)N

(Model No. 26230, No. 26230-RS, No. 26235, No. 26235-RS
only - marked “Version 4”)

Printer

Paper Roll Length:

Approximately 80 feet (960")

Tests/Roll:

Versions 1 and 2:

Approximately 420 Tymps/Reflex or 210 people

Versions 3 and 4:

Approximately 230 tests or 115 people

Assumption:

2 Tymps/Reflex + 1 Audiogram per person)

Speed:

Approximately 1.5 minutes to print three screens:

Tympanogram

Tympanogram + reflex (4)

Audiogram