sChar

Slow Index

The data index where the last compliance data point
measured at 600 daPa/sec before the rate changes to
200 daPa/sec is stored.

-1 = No rate change occurred

33

2

sChar

Fast Index

The data index where the first compliance data
point measured at 600 daPa/sec after changing back
to 600 daPa/sec from the 200 daPa/sec rate is
stored.

-1 = No rate change back occurred

35

2

uChar Tymp Data [0]

Tympanometry compliance data point #0 in

cm

3

,

stored as: compliance X 64. Range = 0.00 to 3.98

cm

3

.

A maximum of 88 data points are stored per
Tympanometry test.

:

:

:

:

:

209

2

uChar Tymp Data [87] Tympanometry compliance data point #87 in

cm

3

,

stored as: compliance X 64

211

2

uChar Tymp Scale

Tympanometry compliance axis scale. 15 = 1.5

cm

3

30 = 3.0

cm

3

213

2

uChar Number of

Reflex Tests

The number of reflex tests performed Range = 0 to
4

215

2

uChar Reflex Test

Parameters

Reflex test selection parameters.

Bit 0 = Ispi status:

0 = Not selected

1 = Selected

Bit 1 = Contra status:

0 = Not selected

1 = Selected

Bit 2 = 500 Hz status:

0 = Not selected

1 = Selected

Bit 3 = 1000 Hz status:

0 = Not selected

1 = Selected

Bit 4 = 2000 Hz status:

0 = Not selected

1 = Selected

Bit 5 = 4000 Hz status:

0 = Not selected

1 = Selected

Bits 6 & 7 = Not used

Character

Number

Number of

Characters

Data
Type

Field Name

Field Description