Audiometry test results record, Notes, Audiometry test results record -9 notes -9 – Welch Allyn TM 262 Auto Tymp - User Manual User Manual
Page 75: Rs-232 interface
Revised 4/11/08
5-9
RS-232 Interface
Audiometry test results record
Notes
1.
“uChar” & “sChar” designate unsigned and signed characters respectively, single bytes
represented in Hexadecimal by two ASCII characters.
Example: OxE9 is sent as: “E”, “9”
2.
“ulnt: and slnt” designate unsigned and signed 16-bit integers respectively, expressed in
HiByte/ LowByte sequence by four Hex ASCII characters.
Example: OxE196 is sent as: “E”, “1', “9”, “6”
3.
Tympanometry compliance values are stored in the record scaled by 64. To convert to cm
3
,
divide by 64.
Example: Tymp Data[0] = “4”, “2” = 0x42=66 decimal scales X64 = 66/64 = 1.0-3 cm
3
4.
Reflex compliance values are stored in the record scaled by 256: To convert to cm
3
, divide
by 256.
Character
Number
Number of
Characters
Data Type
Field Name
Field Description
1
1
ASCII
Start of record
“.”
2
1
ASCII
Record Type
“y”
3
1
ASCII
Record Sequence Number
“0” to “9”
4
1
ASCII
Record Index Number
“1” to “8”
5
1
ASCII
Total Record Number
“1” to “8”
6
11
13
5
2
4
ASCII
uChar
sint
Reserved Ear
HL Threshold 125 Hz
Either the RIGHT ear
is selected OR the
LEFT ear is selected.
Both ears selected or
no ears elected is
invalid.
-10 to +100 dB HL x 2
NR = Any value in the
range of 231 to 450
NT = 32,768 (0x8000
Hexadecimal)
:
:
:
:
:
53
4
sint
HL Threshold 8000 Hz
-10 to +100 dB HL x 2
NR = Any value in the
range of 231 to 450
NT = 32,768 (0x8000
Hexadecimal)
57
2
uChar
Checksum
The hexadecimal sum
of characters 1 to 56
59
2
ASCII
Package Terminator
“CR”, “LF”