Audiometry test results record, Notes, Audiometry test results record -9 notes -9 – Welch Allyn TM 262 Auto Tymp - User Manual User Manual

Page 75: Rs-232 interface

background image

Revised 4/11/08

5-9

RS-232 Interface

Audiometry test results record

Notes

1.

“uChar” & “sChar” designate unsigned and signed characters respectively, single bytes
represented in Hexadecimal by two ASCII characters.

Example: OxE9 is sent as: “E”, “9”

2.

“ulnt: and slnt” designate unsigned and signed 16-bit integers respectively, expressed in
HiByte/ LowByte sequence by four Hex ASCII characters.

Example: OxE196 is sent as: “E”, “1', “9”, “6”

3.

Tympanometry compliance values are stored in the record scaled by 64. To convert to cm

3

,

divide by 64.

Example: Tymp Data[0] = “4”, “2” = 0x42=66 decimal scales X64 = 66/64 = 1.0-3 cm

3

4.

Reflex compliance values are stored in the record scaled by 256: To convert to cm

3

, divide

by 256.

Character
Number

Number of
Characters

Data Type

Field Name

Field Description

1

1

ASCII

Start of record

“.”

2

1

ASCII

Record Type

“y”

3

1

ASCII

Record Sequence Number

“0” to “9”

4

1

ASCII

Record Index Number

“1” to “8”

5

1

ASCII

Total Record Number

“1” to “8”

6

11

13

5

2

4

ASCII
uChar

sint

Reserved Ear

HL Threshold 125 Hz

Either the RIGHT ear
is selected OR the
LEFT ear is selected.
Both ears selected or
no ears elected is
invalid.

-10 to +100 dB HL x 2

NR = Any value in the
range of 231 to 450

NT = 32,768 (0x8000
Hexadecimal)

:

:

:

:

:

53

4

sint

HL Threshold 8000 Hz

-10 to +100 dB HL x 2
NR = Any value in the
range of 231 to 450

NT = 32,768 (0x8000
Hexadecimal)

57

2

uChar

Checksum

The hexadecimal sum
of characters 1 to 56

59

2

ASCII

Package Terminator

“CR”, “LF”