normally approximately 20,000 counts. To compute the number of counts divide the set
capacity by the increment size.
F6.3 Enter Cal Mode
This is how you calibrate the scale. Enable this and continue on to F6.4. If you do not
enable it you will advance to CAL d (calibration done) and the calibration will be
unchanged and you will exit from set-up mode.
F6.4 Linearity
Allows you to correct for small linearity errors in the middle of the weight range. This is
not allowed in Legal for Trade applications. If not used, this should be set to 0 if set
to -0, linearity will not be affected but you will be required to place full capacity
weight for calibration.
Set-Up Procedure
Place the internal CAL switch (S1-1) to the ON position. The scale must be configured
before operation by programming the needed parameters. Use the following front panel
switches to view and/or change existing data values:
[ZERO] Used to scroll backward through the programming sequence steps.
[TARE] Used to scroll forward through the programming sequence steps when in
function select mode and to sequence through available choices of individual functions.
[PRINT] Used to accept displayed data then to advance one step in the programming
sequence.
[CLEAR] Used to return to the Function Select Enable portion of the SET-UP procedure.
[F1] Used to advance to the end of the SET-UP procedure to save or discard the changes
entered. This button is not labeled and is located below the [RECALL TARE] button.
FSE Function Select Enable Selects the starting function group to be programmed. Use
[ZERO] and [TARE] buttons to move back and forth through this menu. Press print to
enter that section at the first item.
Choices: F1, F2, F3, F4, F5, F6 (F7 on some custom systems)
FSE F1
F1.0 Raw Counts Display Enable
d - Show normal counts on display.
Page 26