Archived 4/2/10 – ETS-Lindgren HI-1710 Microwave Measurement System (Archived) User Manual

Page 7

background image

Page — 2

HI

n

1710 Manual

indicator reads with two or three decimal place resolution
from zero to 10 mW/cm

2

. Temporary over-exposure to

microwave fields up to 2000 mW/cm

2

will not cause

probe burn-out or damage the instrument or its
calibration.
Four ranges with full scale indications of 1, 2, 5, and 10
mW/cm

2

or auto-range can be selected from the front

panel. With the instrument set to auto-range, the
microprocessor controller will select the best range for
each measuring condition. Selecting the range,
determines the full scale setting of the bar graph display
located below the digital readout to give a continuous
analog indication of the leakage level being measured.
The audio alarm circuit can be set to any level up to 10
mW/cm

2

. Whenever the level of the microwave leakage

exceeds the preselected value, the audio alarm will
sound, alerting the operator that the chosen level has
been exceeded.