In this message, XXXX is the raw count output from the
A/D converter and the messages are:

X VOLT SUPPLY TOO LOW "x" is 4, 5, or 8
X VOLT SUPPLY TOO HIGH "x" is 4, 5, or 8
NEG. 8 VOLT SUPPLY TOO LOW
NEG. 8 VOLT SUPPLY TOO HIGH
OFFSET VOLTAGE TOO LOW
OFFSET VOLTAGE TOO HIGH
BIAS VOLTAGE TOO LOW
BIAS VOLTAGE TOO HIGH
CANNOT CAL. LOW ANALOG OUTPUT
CANNOT CAL. HIGH ANALOG OUTPUT
STACK OVERFLOW
MISSED INTERRUPT INTERVAL
STIM TEST TOO LOW
STIM TEST TOO HIGH
TIMEOUT WAITING FOR CHARACTER

Refer to section 6.0 (Test and Troubleshooting) for
descriptions of the error codes displayed on the digital
display.

Connection and Operation of the RS-232 Serial Data Port

Refer to Appendix A for a listing of the HI-1710 serial
data port commands.

Connect the device as shown in the appendices. The
HI-1710 serial data port is configured for eight (8) data
bits and one (1) stop bit.

Commands are sent to the HI-1710 using capital letters
and numbers with no spaces. A command is completed
with a Carriage Return (CR). The HI-1710 has a 32
character buffer, therefore, groups of commands can be
sent, eg., a complete test setup. When sending a group