Section 13
Standard Configuration
Surface
Ensure that the area for the instrument is a solid, level surface that can
support the weight of the system (approximately 240 lbs.)
Size
Minimum work area dimensions: 2.0 m W x 0.8 m D x 1.0 m. H
Environment The work area must have an ambient temperature between 18°C –
26°C (64°F – 79°F), not facing direct sunlight and not effected by a
hood or vent.
Power
Source
Dako recommends a dedicated power source to prevent interference
from other instruments or equipment.
Reagent Probe Wash
Station
X Axis
Mechanism
Z Head
Assembly
Back
Reagent Rack b
Right
Reagent Rack a
Left
Sink
(Beneath slide racks)
Y Axis
Mechanism
Front
Slide Rack
Slide Rack
Handle
Glass Slide
Autostainer Handbook