Maintenance and troubleshooting, Troubleshooting – Dako Autostainer User Manual
Page 45

Section 11
Maintenance and Troubleshooting
 
The Autostainer should be cleaned after 150 slides have been processed or after one 
week, whichever comes first. The instrument keeps track of the number of staining runs 
performed and displays a maintenance message listing the number of slides since the 
last cleaning cycle. 
 
Note: Please contact your Dako representative regarding maintenance details. 
Troubleshooting
Problem
Possible Cause
Course of Action
Autostainer will not turn on (top 
green LED not on). Message 
displays “Instrument Not On” when 
a run is started. 
 
The Autostainer is 
unplugged at the power 
source. 
 
Check the power cord 
connections and verify the 
surge protector is plugged in 
and turned on. 
 
The Autostainer does not respond 
when you initiate a prime pump or 
a run (bottom LED not on). 
Message displays “Instrument Not 
On” when a run is started. 
The cable between the 
Autostainer and the 
computer is not connected 
properly. 
 
Shut down the system and 
turn the power off. Disconnect 
and reconnect the cable 
between the Autostainer and 
the computer. 
Please use caution. Pins 
can be bent and cause 
damage to the instrument 
during operation. 
 
Autostainer suddenly stops during 
a run. 
 
The Autostainer will stop 
frequently during the run. 
This is due to set 
incubation times. Check 
the run log to verify. 
 
No action is needed. The 
Autostainer will resume 
movement. Check Run Log 
screen. 
Buffer or DI Water does not flow 
out during a run or a prime. 
 
Autostainer was not primed 
properly. 
Prime pump. Pump may need 
to be activated several times. 
Not enough buffer or DI 
water in the appropriate 
reservoir. 
 
Fill the reservoir with the 
proper level of fluid. Then 
prime the pump. 
 
Bent or pinched tubing. 
 
Straighten out the tubing and 
prime the pump. 
 
 
 
 
Autostainer Handbook
37
