Dako Autostainer User Manual
Page 48

Problem
Possible Cause
Course of Action
Slide number is 
not listed on 
label. 
Label is not 
centered correctly 
on label. 
Reprint label if necessary and place the label 1/2 cm 
from the top and 1/8 cm from the side of the slide. 
Computer 
functions lock up 
during a run, but 
the Autostainer is 
still performing 
the run. 
Windows error.
Allow the Autostainer to finish the run. Reboot system 
before the next run. 
User is unable to 
access many of 
the functions in 
the Initialize 
section of the 
program. 
User has been set 
up as a “Non-
Authorized” user. 
Contact a Dako representative.
Illegal or fatal 
exception errors. 
Windows error.
Shut down the system and turn the power off at the 
surge protector. Wait 30 seconds and then power up 
the system. If this problem persists, contact a Dako 
representative. 
EWP/DWP 
messages on 
Run Log. 
Enable waste 
pump/disable 
waste pump. 
Waste is not being pumped out of the system within 
the expected time frame. If waste is still separating, 
contact a Dako Technical support for additional 
cleaning instructions. If waste is no longer separating, 
contact Dako for service. 
EWL/PFR 
messages on 
Run Log. 
Faulty waste 
removal/probe 
failed to retract. 
Contact a Dako representative.
The computer 
turns on but 
“Safe Mode” is 
displayed on all 
four corners of 
the screen. 
The computer has 
been improperly 
shut down. 
Shut down the computer by using the Windows Start 
button located at the bottom left corner of the screen 
and selecting Shut Down. This button may be hidden 
requiring you to drag the mouse pointer off the bottom 
of the screen for it to appear. Always shut down the 
computer using this method. Perform Scan Disk 
through Windows Explorer/Accessories/System 
Tools/Scan Disk. For assistance, contact a Dako 
representative. 
Monitor screen 
went blank 
(power LED is 
green). 
Screen saver utility 
installed. 
Move the mouse. Current screen will reappear. 
Remove screen saver as this may interfere with the 
Autostainer operation. 
Monitor screen 
went blank
(power LED is 
orange). 
The cable between 
the monitor and the 
computer is not 
connected 
properly. 
Disconnect and reconnect the cable between the 
monitor and the computer. 
 
 
 
 
 
 
 
 
40
Autostainer Handbook