Dako Autostainer User Manual
Page 47

Problem
Possible Cause
Course of Action
The probe and 
wash/blow head 
are not going to the 
correct locations. 
The arm has been 
knocked out of 
calibration. 
Abort run and save run log. Reprogram the run and 
delete the steps from the Protocol Design screen 
that have already been completed. Restart run. 
Bent probe tip.
Verify that all reagent vial caps have been removed 
and slide racks are firmly locked in place in the 
down position. Contact a Dako representative. 
No Staining
Insufficient reagent 
in vial. 
If the system detects insufficient reagent to 
complete the current step, you have one minute to 
add reagent, after which some slides may be 
skipped. 
 
Reagents vials were 
not loaded in the 
correct locations in 
the reagent racks. 
Check the Reagent Map to verify proper locations. 
Check Run Log for “Unable to Locate Reagent” 
Skipped Slides message. 
Slides were not 
loaded in the 
correct locations. 
Check the Load Slide printout to verify proper 
locations. 
Slides were loaded 
with tissue side 
down. 
Orient slides properly.
Programming error.
Check the Programming Grid to ensure that the 
staining run was programmed correctly. 
Slides drying out.
Make sure adequate dispense volume that is 
suitable for desired incubation times is programmed. 
Insure that no extreme environmental conditions are 
contributing to the drying of the slides (i.e. direct 
sunlight, heating/air conditioning vents, hoods, etc.) 
The text of the label 
does not begin 
along the left edge 
of the label and the 
text is getting cut 
off. 
The Seymour 
printer may need to 
be reset. 
To reset the Seymour Label Printer, turn off the 
printer. Hold down the green label feed button on 
the top of the printer. Turn on the printer while 
holding down the green button and continue holding 
the button for three seconds. The labels advance 
and the printer will print. Release the feed button. 
Press and release the feed button one more time to 
advance one label. Perform a test print. 
Error occurs during 
printing. 
Labels or ribbon 
have run out. 
Replace ribbon in the printer and the printer will 
restart. 
Incorrect COM port 
selected. 
Select the correct COM port from the Slide Label 
Design screen, see the Initializing section. 
Incorrect printer is 
selected. 
Select the correct printer model from Slide Label 
Design screen, see the Initializing section. 
 
 
 
 
 
 
 
 
 
 
Autostainer Handbook
39