Displaying of the variables – Carlo Gavazzi WM4-96 User Manual
Page 16

▲
▲
16
Displaying of the variables
Inputs
11
Energy Meters pages
8
18
❑
Pass "998": display of the stored dmd variables
(Only if the RS232 + RTC module is present)
The stored dmd variables can be displayed by pressing
the
key in any one of the measuring page (excluded the
“load profile” pages): when the instrument asks for the
, enter the value “998”: the instrument will
display the values of the stored variables, the time (hours-
minutes-seconds) and the date (day-month) of storage.
Use
the and keys
to
scroll
all the values; to exit press the
key. If the value has not been
stored, the instrument displays
“NOT AVAILABLE VALUE”.
❑
Pass "997": Reset of the Data Memory (only if the
RS232+RTC module is available)
It’s possible to reset all the information stored in the data mem-
ory. To enter this function press the
key in any one of the
measuring pages (excluded the “load profile” pages), when the
instrument asks for
, enter the value “997” and
confirm with
, then press
to confirm the reset; press
not to execute the reset.
❑
Pass "996": Reset of the load profiles
It’s possible to reset all the information relating to the
load profiles. To enter this function, press the
key in any
one of the measuring pages (excluded the “load profile
pages”), when the instrument asks for
enter the value “996” and confirm with
, then press
again
to confirm the reset, press
not to execute the reset.