14 injection valve, 1 connecting the injection valve, Connecting the injection valve – Metrohm 850 Professional IC Anion MCS Prep 2 User Manual
Page 54: Figure 24, Injection valve – connected, See chapter 2.14, For sample preparation. see chapter 2.14
2.14 Injection valve
■■■■■■■■■■■■■■■■■■■■■■
44
■■■■■■■■
850 Professional IC – Anion – MCS – Prep 2
4 Guide the other end of the connection capillary out of the instrument
through a capillary feed-through.
Caution
If the sample degasser is not used, the input and output must be
sealed with the 6.2744.220 thread stoppers.
2.14
Injection valve
The injection valve connects the eluent and sample path. Through rapid
and precise valve switchover a precise amount of sample solution defined
by the size of the sample loop is injected and rinsed with eluent onto the
separation column.
2.14.1
Connecting the injection valve
The injection valve has six connectors: two for the sample path (connec-
tors 1 and 2), two for the eluent path (connectors 4 and 5) and two for
the sample loop (connectors 3 and 6).
Note
The capillaries of the eluent path and the sample path and the sample
loop are already installed in the newly delivered instrument.
1
2
3
4
5
6
1
2
3
4
5
6
7
7
7
7
Figure 24
Injection valve – connected
1
Injection valve
2
Sample loop
Connected to connectors 3 and 6.