5 troubleshooting, 1 problems and their solutions, Problems and their solutions – Metrohm 850 Professional IC Anion MCS Prep 2 User Manual
Page 113
■■■■■■■■■■■■■■■■■■■■■■
5 Troubleshooting
850 Professional IC – Anion – MCS – Prep 2
■■■■■■■■
103
5 Troubleshooting
5.1
Problems and their solutions
Problem
Cause
Remedy
Marked drop in
pressure
Leak in the system.
Check and seal connections (see Chapter 2.5,
page 17).
Marked rise in pres-
sure
6.2821.120 inline filter
blocked.
Replace 6.2821.130 filter (see Chapter 4.6,
page 86).
MSM – blocked.
■
Regenerate the MSM (see Chapter
4.11.3.1, page 91).
Note: 6.2821.180 pump tubing connection
with filter must be used (30-
Conductivity detector
blocked.
■
Shorten capillary ends by several mm (see
Chapter 4.14.1, Page 100).
■
Rinse detector against the normal flow
direction (see Chapter 4.14.1, Page 100).
Guard column – blocked.
Replace guard column (see Chapter 2.21,
page 65).
Separation column –
blocked.
■
Regenerate separation column (see Chap-
ter 4.15.4, page 101).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 68).
Note: Samples should always be microfiltered
(see Chapter 4.7, page 88).
Injection valve – valve
blocked.
Have the valve cleaned (by Metrohm service
technicians).
Drift of the baseline
Thermal equilibrium not yet
attained.
Condition (see Chapter 2.15, page 47) instru-
ment with the column thermostat switched
on .
Leak in the system.
Check and seal connections (see Chapter 2.5,
page 17).