11 inline filter, See chapter 2.11 – Metrohm 850 Professional IC Anion MCS Prep 2 User Manual
Page 50
2.11 Inline filter
■■■■■■■■■■■■■■■■■■■■■■
40
■■■■■■■■
850 Professional IC – Anion – MCS – Prep 2
4 Setting the flow rate
■
Start MagIC Net (if not yet started).
■
Ensure that the eluent aspiration tubing is immersed sufficiently in
the eluent.
■
Let the high pressure pump run.
5 Aspirating eluent
■
Aspirate with the syringe (20-1) until bubble-free eluent flows into
the syringe.
6 Completing deaerating
■
Turn off high pressure pump.
■
■
Remove syringe (20-1) from the Luer connector (20-2).
■
Pull the purging needle (20-3) out of the deaerating capillary
(20-4).
2.11
Inline filter
Between the purge valve and the pulsation damper the 6.2821.120 inline
filter is installed as protection against particles.
Inline filters protect the separation column securely against possible con-
tamination from the eluent. Inline filters can however also just as well be
used for the purpose of protecting the suppressor against contaminations
in the regeneration or rinsing solutions. The fine 2 µm material of the
readily and easily replaceable filter platelets removes particles such as bac-
teria and algae from the solutions.
Note
The inline filter is already installed in the newly delivered instrument.
The following installation instructions need not be carried out at the
time of initial installation.