beautypg.com

13 metrohm co2 suppressor (mcs), 14 conductivity measuring system, 13 metrohm co – Metrohm 850 Professional IC Anion MCS Prep 2 User Manual

Page 122: Suppressor (mcs)

background image

6.13 Metrohm CO

2

suppressor (MCS)

■■■■■■■■■■■■■■■■■■■■■■

112

■■■■■■■■

850 Professional IC – Anion – MCS – Prep 2

6.13

Metrohm CO

2

suppressor (MCS)

Material

fluoropolymer

Resistance to sol-
vents

No restriction (apart from PFC)

Vacuum

Working range

Microprocessor-controlled / stabilized

Build-up time
after start

< 30 s

Capillary volume

400 µL

Recommended
flow range

0.1…1.0 mL

6.14

Conductivity measuring system

Type

Microprocessor-controlled Digital Signal Processing (DSP technol-
ogy)

Intelligent detector with 6 reference chromatograms

Measuring range

0…15000 µS/cm single range

Noise

< 0.1 nS at 1 µS/cm

Deviations from
the linearity

< 1 % for conductivity values greater than 1…16 µS/cm (typical for
analyses with sequential suppression)

Drift

< 0.2 nS/cm per hour

Measuring rate

10 measurements per second for optimum results without filtering

Resolution

0.0047 nS/cm

Baseline

Noise < 0.2 nS/cm typical for sequential suppression

Conductivity
detector

Cell volume

0.8 µL

Cell constant

Individual calibration data saved in the detector

adjustable range: 13.0…21.0 /cm

Electrodes

Ring-shaped electrodes made from stainless steel

Materials in
contact with
eluent

Chemically inert PCTFE

Maximum oper-
ating pressure

5.0 MPa (50 bar)