13 metrohm co2 suppressor (mcs), 14 conductivity measuring system, 13 metrohm co – Metrohm 850 Professional IC Anion MCS Prep 2 User Manual
Page 122: Suppressor (mcs)
6.13 Metrohm CO
2
suppressor (MCS)
■■■■■■■■■■■■■■■■■■■■■■
112
■■■■■■■■
850 Professional IC – Anion – MCS – Prep 2
6.13
Metrohm CO
2
suppressor (MCS)
Material
fluoropolymer
Resistance to sol-
vents
No restriction (apart from PFC)
Vacuum
Working range
Microprocessor-controlled / stabilized
Build-up time
after start
< 30 s
Capillary volume
400 µL
Recommended
flow range
0.1…1.0 mL
6.14
Conductivity measuring system
Type
■
Microprocessor-controlled Digital Signal Processing (DSP technol-
ogy)
■
Intelligent detector with 6 reference chromatograms
Measuring range
0…15000 µS/cm single range
Noise
< 0.1 nS at 1 µS/cm
Deviations from
the linearity
< 1 % for conductivity values greater than 1…16 µS/cm (typical for
analyses with sequential suppression)
Drift
< 0.2 nS/cm per hour
Measuring rate
10 measurements per second for optimum results without filtering
Resolution
0.0047 nS/cm
Baseline
Noise < 0.2 nS/cm typical for sequential suppression
Conductivity
detector
Cell volume
0.8 µL
Cell constant
■
Individual calibration data saved in the detector
■
adjustable range: 13.0…21.0 /cm
Electrodes
Ring-shaped electrodes made from stainless steel
Materials in
contact with
eluent
Chemically inert PCTFE
Maximum oper-
ating pressure
5.0 MPa (50 bar)