Bird 5011, 5015, 5011-EF and 5015-EF Terminating Sensors
General
Corrected terminated average power
Sensor Type
Diode based terminated true average
power measurement
Frequency Range
5011
5011-EF
40 MHz to 4 GHz
40 MHz to 12 GHz
RF Power Range
10
m
W to 10 mW
(–20 dBm to +10 dBm)
Max Power
(Damage Level)
2 W avg.
125 W peak for 5 μs
Peak/Average Ratio
12 dB max
Uncertainty
*
*. 5011: Above 40 °C, when making measurements at
frequencies between 40 and 100 MHz, add 1%. 5011-
EF: Above 40 °C or below 10 °C, add 1%.
± (5% of reading ± 1 μW) (95% c.l.)
(excluding mismatch uncertainty)
(with correction factors for -EF)
RF Connector
5011
5011-EF
N Male
Precision N Male
Impedance, Nominal
50 ohms
Input VSWR:
5011
Typical
Maximum
5011-EF
Typical
Maximum
1.03 (36.6 dB return loss)
1.20 (20.8 dB return loss)
1.05 (32.0 dB return loss)
1.25 (19.1 dB return loss)
Power Supply
From host instrument via cable
Temp, Operating
–10 to +50 °C (14 to 122 °F)
Temp, Storage
–40 to +80 °C (–40 to +176 °F)
Mechanical Shock
IAW MIL-PRF-28800F class 3
Vibration
IAW MIL-PRF-28800F class 3
Humidity, Max
95% (non-condensing)
Altitude, Max
4,500 m (15,000 ft.)
CE
CE compliant
ROHSS
ROHSS compliant
Dimensions, Nominal 6” long x 1.5” diameter
Weight, Max
0.75 lb. (0.35 kg)
Recommended
Calibration Interval
12 Months