Bird 5011, 5015, 5011-EF and 5015-EF Terminating Sensors

General

Corrected terminated average power

Sensor Type

Diode based terminated true average
power measurement

Frequency Range

5011
5011-EF

40 MHz to 4 GHz
40 MHz to 12 GHz

RF Power Range

10

m

W to 10 mW

(–20 dBm to +10 dBm)

Max Power
(Damage Level)

2 W avg.
125 W peak for 5 μs

Peak/Average Ratio

12 dB max

Uncertainty

*

*. 5011: Above 40 °C, when making measurements at

frequencies between 40 and 100 MHz, add 1%. 5011-
EF: Above 40 °C or below 10 °C, add 1%.

± (5% of reading ± 1 μW) (95% c.l.)
(excluding mismatch uncertainty)
(with correction factors for -EF)

RF Connector

5011
5011-EF

N Male
Precision N Male

Impedance, Nominal

50 ohms

Input VSWR:
5011

Typical
Maximum

5011-EF

Typical
Maximum

1.03 (36.6 dB return loss)
1.20 (20.8 dB return loss)

1.05 (32.0 dB return loss)
1.25 (19.1 dB return loss)

Power Supply

From host instrument via cable

Temp, Operating

–10 to +50 °C (14 to 122 °F)

Temp, Storage

–40 to +80 °C (–40 to +176 °F)

Mechanical Shock

IAW MIL-PRF-28800F class 3

Vibration

IAW MIL-PRF-28800F class 3

Humidity, Max

95% (non-condensing)

Altitude, Max

4,500 m (15,000 ft.)

CE

CE compliant

ROHSS

ROHSS compliant

Dimensions, Nominal 6” long x 1.5” diameter
Weight, Max

0.75 lb. (0.35 kg)

Recommended
Calibration Interval

12 Months