5000XT DPM Quick Start Guide

Installing the Battery

1. Lay the DPM, display side down, on a clean surface.
2. Remove the two mounting screws from the back cover.
3. Remove the back cover.
4. Install the battery.
5. Replace the back cover.
6. Secure the back cover with the mounting screws.
7. Attach the DPM to an external power source to charge the battery.
8. Charge the battery for five hours before use.

Connecting

Note:

If the 5000-XT is connected to a PC when starting or restarting the

PC, it may cause the boot process to lock up. Disconnect the 5000-XT before

starting or restarting the PC.

Note:

Use the appropriate Bird Power sensor for the RF power measurement

application. An attenuator may be required.

WARNING

Do not connect to any external power before disassembly.

The potential for electric shock exists.

WARNING

Never attempt to connect or disconnect RF equipment from the transmission line

while RF power is being applied.

Leaking RF energy is a potential health hazard.

RF voltage may be present in RF element socket.

Keep element in socket during operation.

Back Cover

Mounting
Screws

Battery