beautypg.com

Installation 8, 3 installation instructions – Rockwell Automation 836E Solid-State Pressure Sensors User Manual

Page 8

background image

Installation

8

3. 3

Installation Instructions

Fig. 3: Typical applications of the Bulletin 836E to measure pressure in gases, steam and liquids.

Bulletin 836E pressure switch

Shut-off assembly

U-pipe

For typical applications of the Bulletin 836E, see Figure 3 above:
• The image on the left illustrates how the 836E should be applied for pressure measurement in

gases. The switch should be mounted with a shut-off assembly above the sampling nozzle so
that any condensate can drain off into the process.

• The center image shows the correct installation for pressure measurement in steam. Note how

the Bulletin 836E is mounted with a U-pipe below the sampling nozzle. Fill the U-pipe with fill
fluid before commissioning.

• For pressure measurement in liquids, see the drawing on the right of Figure 3 above. The

Bulletin 836E should be mounted below or at the same level as the sampling nozzle.

Additional considerations:
• Do not mount the device in the product flow or at a point where it could be affected by

pressure pulses.

• Calibration and testing are easier if the device is mounted downstream of a shut-off assembly.
• The orientation of the Bulletin 836E can result in zero point shift, i.e. the measured value does

not display zero in a non-pressurized state. This zero point shift can be corrected (see Section 4
– Operation).

• The digital display can be electronically rotated 180° (see Section 4 – Operation).
• The housing can be rotated up to 310° for optimal readability and ease of wiring.

Gas

Steam

Liquid