MTS SWIFT 50 GLP Sensor User Manual
Page 25

Specifications
SWIFT 50 GLP Sensors Installation
Hardware Overview
25
Transducer Interface
Parameter
Specification
Physical
Height
Width
Depth
Weight
Rack Mounting Kit
31.75 mm (1.25 in)
431.8 mm (17 in)
215.9 mm (8.5 in.)
*
1.68 kg (3 lb 11.1 oz)
Optional
*
Add 25.4 mm (1.0 in) for ground lugs.
Environmental
Ambient temperature
Relative humidity
0° C (32° F) to 50° C (122° F)
0 to 85%, non-condensing
Power Requirements
Input voltage
Supply current
Fuse
10–17 V DC or 18–34 V DC
†
2 A typical, 3 A maximum at 12 V DC
3 A fast-blow
† Specified at time of purchase.
Angular velocity
Encoder limit
Processing limit
Encoder resolution
2,200 rpm maximum
10,000 rpm maximum
2048 counts per revolution
(512 pulses with quadrature)
Time delay (encoder tick to main output stable)
12 µs (typical)
Transducer cable length
100 ft maximum
Shunt cable length
100 ft maximum
Analog outputs
Voltage
Capacitive load
Current
Noise at output, with typical gains
±10 V range
‡
(force and moment outputs)
0–5 V sawtooth (angle output)
0.01 µF maximum
6 mA maximum
7 mVpp, DC - 500 Hz (typical)
15 mVpp, DC - 500 Hz (maximum)
‡ Standard from MTS. Other full scale voltages can be evaluated and may be provided at special
request.
Bandwidth (bridge input to main output)
–3 dB at 30.1 kHz (typical)
90 degree at 16.6 kHz (typical)