INFICON STM-2 USB Thin Film Rate/Thickness Monitor User Manual
Page 77
5 - 9
PN
07
4-
61
3-
P1
A
STM-2 Operating Manual
Zeroing commands
B
None
B
Zeroes timer and thickness
C
None
C
Zeroes thickness
D
None
D
Zeroes timer
Film commands
= nnn to update parameter
? to query parameter
q
=[8 char string],[?]
q=COPPER
Sets the current film name to COPPER
E
=[0.500 - 99.99],[?]
E=1.23
Set current film density to 1.23
F
=[0.100 - 9.999],[?]
F?
1.23
Returns current film Z factor
J
=[10.0 - 399],[?]
J = 80.1
Set current film tooling to 80.1%
r
=[1 – 50],[?]
r=10
Sets number of samples to 10 for temporal
averaging
Runtime data commands
M
None
M
!
Return crystal fail status, was failed
@ =
crystal good
!
=
crystal failed
S
None
S
-1595
Return thickness value -1.595 kÅ
s
None
s
-1.6
Return film mass -1.6 µg/cm²
T
None
T
12.4
Return rate, 12.4 Å/s
t
None
t
23.4
Return mass accumulation rate in µg*s/cm²,
23.4 µg*s/cm²
U
None
U
5319234.34 Return sensor frequency 5,319,234.34Hz
V
None
V
12.4
Return crystal life, 12.4% remaining
W
None
W
12:45
Return timer, 12 min 45 s
a
None
a
A
Return RESET Status, true
@ =
OK
A
=
lost power
D
=
lost NONV memory
E
=
lost power and NONV memory
NOTE:
NONV refers to non-volatile memory,
i.e. memory that is persistent even
when power is lost. This status is
indicating that there was a loss of
data, either due an internal database
problem, or a database purge.
Table 5-2 STM-2 command set (continued)
STM-2
command
Parameter
Send
Returned
Description