2 display settings – INFICON STM-2 USB Thin Film Rate/Thickness Monitor User Manual
Page 42
3 - 16
PN
07
4-
61
3-
P1
A
STM-2 Operating Manual
3.3.2.2 Display Settings
Click to display a window to configure the settings for Rate Resolution, Filtering,
Format, and View of the STM-2 window (see
).
Figure 3-11 Display Settings window
Rate Resolution . . . . . . . . . . . . . . . Use the drop-down list box to select 0.0 or
0.00 as the rate resolution to be displayed on
the STM-2 window.
Filtering . . . . . . . . . . . . . . . . . . . . . . 1 to 99
Use the spin box to enter the number of
samples to be averaged together for the
STM-2 window Rate display.
NOTE: This value will not affect graphs or
data logs. Use the Samples
Averaged parameter (see
) for averaging to affect all
displays and data logs.
Format . . . . . . . . . . . . . . . . . . . . . . . Select Thickness in kÅ, Thickness in Å, or
Mass in µg/cm
2
to be displayed on the
STM-2 window and on the Thickness graph.
NOTE: Mass in µg/cm
2
will display Rate as
a Mass Rate in ng/cm
2
/s.
View/Edit . . . . . . . . . . . . . . . . . . . . . Select Application On Top to display the
STM-2 window in front of all other open
windows. Select Lock Parameters to
disable editing of Density, Z-Ratio, and
Tooling on the STM-2 window.