beautypg.com

Ordering information, Pointwatch eclipse co2 detector, Calibration equipment – Det-Tronics PIRECL Infrared Carbon Dioxide Gas Detector PointWatch Eclipse User Manual

Page 26: Spare parts, Assistance

background image

24

95-8676

2.1

orDering informAtion

poInTWaTCh eClIpSe Co

2

DeTeCTor

When ordering, please refer to the PIRECL CO

2

Model

Matrix.

CalIbraTIon eqUIpMenT
The Eclipse CO

2

calibration kits consist of a sturdy

carrying case containing two 3.6 cubic foot (103 liter)

cylinders of specified gas (1x N

2

and 1x CO

2

), a regulator

with a pressure indicator, and three feet of tubing.

Kit, Carbon Dioxide 1%/vol (10,000 ppm)

and Pure Nitrogen

006468-017

PointWatch Eclipse Regulator

162552-002

Eclipse Calibration Bag

006672-002

Spare parTS
Carbon Dioxide, 1%/vol

227117-020

Pure Nitrogen

227117-022

Calibration Port Cover

009192-001

Calibration Magnet

102740-002

Silicone Free Grease

005003-001

O-Ring, 3.75” i.d., for wiring

compartment cover

107427-040

O-Ring, 3.25” i.d., for front

flange (internal)

107427-053

O-Ring, 2.44” i.d., for weather baffle

107427-052

aSSISTanCe
For assistance in ordering a system to meet the needs of

a specific application, contact:

Detector Electronics Corporation

6901 West 110th Street

Minneapolis, Minnesota 55438 USA

Operator: (952) 941-5665 or (800) 765-FIRE

Customer Service: (952) 946-6491

Fax: (952) 829-8750

Web site: www.det-tronics.com

E-mail: [email protected]

table 7—troubleshooting Guide

fault Condition

Corrective action

Low 24 volts

24 vdc operating voltage is out of range. Verify proper wiring to the detector and correct

voltage output from the power source. Power supply faults are self-clearing when the

condition is corrected. If the fault does not clear, consult the factory.

Dirty Optics

Perform cleaning procedure, then re-calibrate as required. (Refer to “Maintenance” for

details.)

Calibration Fault

If the calibration process is allowed to time-out, the fault is set and can only be reset

with a successful calibration. Check the gas bottle to ensure that there is enough gas

to complete the calibration. Are conditions too windy for a successful calibration? If

so, use a PointWatch Eclipse Calibration Bag (P/N 006672-002). Always calibrate

with a Det-Tronics calibration kit for Eclipse with correct regulator. Be sure that the

calibration gas being used matches the configured setting. If the fault is still present,

perform cleaning procedure, then re-calibrate.

Negative Signal Output

This fault is indicated when the signal output drops below 10% of scale. Normally detection

capability is not compromised in this condition. The device was probably zero calibrated with

background gas present. If the condition persists, purge with pure nitrogen and repeat the

zero calibration.

Calibration line active at start-up

The only way to clear this fault is to correct the wiring and reapply power. Be sure that the

calibration line is not shorted and that the calibration switch is open. If the fault does not clear,

consult the factory.

Other Faults

Consult the factory.