beautypg.com

Calibration, Calibration overview – Det-Tronics PIRECL Infrared Carbon Dioxide Gas Detector PointWatch Eclipse User Manual

Page 21

background image

19

95-8676

2.1

cAlibrAtion

CalIbraTIon oVerVIeW
Although routine calibration of the PointWatch Eclipse is

normally not required, the device supports non-intrusive

field calibration capability. Two (2) calibration procedure

options are provided:

1. normal Calibration is a two-step process consisting

of applying nitrogen and mid-scale (span)

adjustment. Calibration gas must be applied by

the operator to enable span adjustment. Prior to

calibration initiation, purge the Eclipse optics with

nitrogen to ensure that an accurate zero condition is

present.

For normal calibration, the recommended calibration

gas concentration is 1%/vol (10,000 ppm).

2. zero only Calibration is a one-step process

consisting of nitrogen (zero) condition adjustment

only, which is performed automatically by the device.

This procedure adjusts the zero signal output only,

and is normally used if the 4 milliampere signal level

has drifted. The cause of drift is typically due to

the presence of background gas during calibration.

Purge the Eclipse optics with pure nitrogen prior

to calibration initiation to ensure an accurate zero

condition is present.

aDDITIonal CalIbraTIon noTeS

Important

Always ensure that the correct gas type is used for
calibration. (2.5 LPM flow rate is recommended.)

NoTe

ensure that the detector has been operating for at
least two hours before calibrating.

NoTe

Always ensure that the eclipse is zero calibrated
with nitrogen.

NoTe

Under very windy conditions, it may not be
possible to successfully calibrate the eclipse. This
situation is easily corrected by using the eclipse
Calibration Bag (P/N 006672-002), available from
Det-Tronics.

NoTe

Always place the protective cap back on
the calibration nozzle after completion of span
calibration.