Setup parameters – Despatch LEB Series User Manual
Page 32

24
Setup Parameters
Note: When changing between
°
C and
°
F, the setup parameters Filt, o1.Hy, HSP.L, and HSP.H
settings convert automatically.
inPt
Input type – this selects thermocouple type. LBB ovens use type J
thermocouple.
unit
Process unit – selects between
°
C and
°
F for reading process
temperature.
rESo
Display resolution – selects the location of the decimal point on process
related parameters.
SHif
PV shift value (offset) – this moves the display temperature to the oven
temperature.
Filt
PV filter – if process value is unstable to read, increasing this value will
steady the input signal.
out1
Output 1 function – this the function of the output. This must be set to .Hi
for
HI-LIMIT control.
o1.Hy Output 1 hysteresis value – the amount of degrees that the temperature
must be below the setpoint temperature before the HI-LIMIT can be reset.
HSP.L Lower limit of HSP1 – the minimum temperature that the HI-LIMIT can be
set.
HSP.H Upper limit of HSP1 – the maximum temperature that the HI-LIMIT can be
Code
Name
Settings
inPt
Input type.
J_tC
unit
Process unit.
°°°°
C (or
°°°°
F)
rESo
Display resolution.
No.dP
SHif
PV shift value (offset).
0
Filt
PV filter.
0
out1
Output 1 function.
Hi.
o1.Hy
Output 1 hysteresis value.
2.0
HSP.L
Lower limit of HSP1.
0 (32 if
°°°°
F)
HSP.H
Upper limit of HSP1.
204 (400 if
°°°°
F)
out2
Output 2 function. (N/A)
None
Addr
Address assignment for digital communication. (N/A)
1
bAud
Baud rate of digital communication. (N/A)
4.8
PAri
Parity bit of digital communication. (N/A)
EVEn
diSP
Normal display format.
PV