Set-up parameters – Despatch LEB Series User Manual
Page 27

19
bS
PV Bias - Expressed in degrees from –199 to 999. This parameter is used
to set the actual oven temperature to the controller display.
LoC Key Lock - This provides levels of access to the controller.
0 = No key lock, full access to controller.
1 = Prevents changing of all parameters except setpoint.
2 = Prevents all parameters from being changed including the setpoint.
-1 = Set to enter the Setup parameter setting display.
The controller is set to LoC=1 at the factory. This prevents inadvertent changing
of control parameters with the exception of setpoint.
If it becomes necessary to alter control parameters, change the LoC=1 to LoC=0.
Follow the parameter programming mode found earlier in this manual.
When LoC = -1, the parameters are displayed in the order shown in the Set-Up
Parameters section, below. Note: If you are unable to change the setup
parameters, go back out and change the LoC = 0. Press the
key until LoC
appears again, change LoC = -1 and press the
key.
Set-Up Parameters
Set-up Parameter Outline
In
Input Type – This parameter is set for type of input and whether PV is in
°
C
or
°
F.
SPH The maximum setpoint limit for oven. The user cannot set the setpoint
above the maximum setpoint.
SPL The minimum setpoint limit of oven. The user cannot change the setpoint
below this lower setpoint limit.
dr
Direct/Reverse action. This value is set for 0 (reverse action).
dSP Priority of PV/SP display. This parameter either displays the PV (process
variable) or SP (setpoint). 0 = process variable, and 1 = setpoint.
Code
Name
Settings
In
Input Type
5 (35 if
°
F)
SPH
Setpoint High
204 (400 if
°
F)
SPL
Setpoint Low
30 (86 if
°
F)
Dr
Direct/Reverse Action
0
DSP
PV/SP Display
0